-200 To -299, Execution Errors - Keysight n9020a Service Manual

Mxa signal analyzer
Hide thumbs Also See for n9020a:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
–221
Settings
conflict;Range <0> is
turned off as total
range points > 40001
–221
Settings conflict;RF
Input 2 is not
available above 1GHz
–221
Settings
conflict;Scan Time &
Points do not apply
in current Step/Time
Control.
–221
Settings
conflict;Step Size &
Meas Time do not
apply in current
Step/Time Control.
–200 to –299, Execution Errors
This section lists all messages in the range except –221 error messages. For
details of –221 messages, see
Note that Execution Errors are divided into subclasses, as follows:
–21x
–22x
–23x
–24x
–25x
–26x
–27x
–28x
–29x
Err#
Message
–200
All ranges are off.
Turn on at least a
range
–200
At Full Zoom
Verbose/Correction Information
Max of Total range points is 400000. Reduce Scan Points or increase Step
Size in order to turn on that range
If the frequency range is set above 1GHz, you cannot change to RF Input 2
The Scan Time & Points are not available when Step/Time Control is set to
Step & Dwell
The Step Size & Meas Time are not available when Step/Time Control is set to
Scan Time&Pts
Trigger errors
Parameter error
Data corrupt or stale (invalid data)
Hardware error
Mass storage error
Expression data error
Macro error
Program error (a downloaded program-related execution error)
Memory use error
Verbose/Correction Information
There are no range turn on in scan table. You need to turn on at least a range to
initiate a scan
Marker Zoom is not available as it has reached full zoom
101
"–221 Settings Conflict Errors" on page
89.