-200 To -299, Execution Errors - Keysight N8973B Service Manual

Nfa series
Hide thumbs Also See for N8973B:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
–221
Settings
conflict;Function not
available while
measurement is
running
–221
Settings
conflict;function
unavailable with this
EMC Standard
–221
Settings conflict;Log
Percent Auto Step
Rule does not apply
in Swept Scan Type.
–221
Settings conflict;QPD
+ EMI Average + RMS
Average is not
allowed
–221
Settings
conflict;Range <0> is
turned off as total
range points > 40001
–221
Settings conflict;RF
Input 2 is not
available above 1GHz
–221
Settings
conflict;Scan Time &
Points do not apply
in current Step/Time
Control.
–221
Settings
conflict;Step Size &
Meas Time do not
apply in current
Step/Time Control.
–200 to –299, Execution Errors
This section lists all messages in the range except –221 error messages. For
details of –221 messages, see
Note that Execution Errors are divided into subclasses, as follows:
–21x
Trigger errors
–22x
Parameter error
96
Verbose/Correction Information
Settings change is not allowed while measurement is running. You must stop
the current measurement if you wish to change the settings
The Log Percent rule is not available when Scan Type=Swept Scan because
we are always doing linear sweep
You cannot turn on all 3 EMI detectors together. You must turn off one of the
EMI Detectors before you turn this on
Max of Total range points is 400000. Reduce Scan Points or increase Step
Size in order to turn on that range
If the frequency range is set above 1GHz, you cannot change to RF Input 2
The Scan Time & Points are not available when Step/Time Control is set to
Step & Dwell
The Step Size & Meas Time are not available when Step/Time Control is set to
Scan Time&Pts
"–221 Settings Conflict Errors" on page
Keysight NFA Series Noise Figure Analyzers Service Guide
85.