4349B
Additional Error Due To the Voltage Source Noise
In the case of capacitor leakage current measurements, the DC voltage source noise aects on
the measurement accuracy. The additional measurement error due to the DC voltage source
noise depends on the DUT's capacitance C
estimations of the additional error when using the DC voltage source that has the following
specications:
External DC Voltage Source Performance
Ripple
Wide band nosie
Switching noise
Figure 8-2. Current Measurement Additional Error (Measurement Time:10 msec)
and resistance R
x
5 Vrms(@50 Hz)
p
1 Vrms/
Hz (@50 Hz)
10 mVrms(@100 Hz)
. Figure 8-2 to Figure 8-5 show
x
General Information
8-9