Preparation For Use; Connection To Power Supply - Siare Falco 202 Evo User Manual

Lung ventilator
Table of Contents

Advertisement

3.2

Preparation for use

3.2.1 Connection to power supply

There are three types of power supplies available on the lung ventilator:
mains power supply (100-240Vac / 50-60Hz)
low voltage power supply (12Vdc / 7A)
battery supply (NiMh 12Vdc - 4,2Ah battery: max. autonomy 4 hours).
Mains power supply
The mains power supply must match the one indicated on the identification plate, placed on
the back of the lung ventilator: 100-240Vac / 50-60Hz.
Insert the power supply cable plug (100-240
Vac) supplied with the device to the plug placed
on the back of the lung ventilator.
Insert the power supply cable plug (100-240
Vac) supplied into the mains socket.
Set the main switch (placed on the back of the
lung ventilator) to "I".
3-4
The electrical connections are a very important part in the installation of the
emergency lung ventilator. Incorrect connections or connections to
unsuitable electrical systems can compromise the safety of the patient and
the User.
Mains power supply must comply with the prescriptions in CEI 64-8/7
standards concerning the locations intended for type A medical use.
WARNING !! Patient / clinician injury hazard
SIARE specialized personnel or qualified technical personnel formerly
authorized by SIARE, shall always verify the following during the
installation:
the presence of the ground cable in the electric plug used to connect
the lung ventilator.
the electric plug used should be equipped with a lock in order to prevent
erroneous placements in wrong plugs without ground.
User manual, DU3104101