Main Characteristics: Anaethesia Unit Structure; Valves Group - Breathing System - Siare Morpheus ND User Manual

Anaesthesia unit
Table of Contents

Advertisement

Standard auxiliary fresh gas outlet
The fresh gas dosing system guarantees maximum safety thanks to the redundant system with
double valve and double control. In this way, the supply of oxygen or overdoses of nitrous oxide in
the event of a failure will never be interrupted.
Emergency back-lit oxygen flowmeter that can be used even when the machine is off.
The incorporated PROTOLOCK system identifies any wrong connection of the incoming gases
during the initial test phase: this device avoids potentially deadly situations for the patient, as already
happened with traditional systems.
1.3

Main characteristics: anaethesia unit structure

The mechanic structure of the anaesthesia unit is made of light aluminium alloy and plastic moulds:
this ensures an excellent impact resistance thanks to its flexibility and excellent abrasion resistance.
The mechanic structure also includes:
the housing for the valves group,
the manual ventilation controls, a mechanical flow meter and three pressure gauges
on the back side are provided the medical gas intakes, which are positioned in a rational and
easily visible way,
two horizontal guide for vaporizer supports (a vaporizer in use and one in parking)
The dimensions and weight of the anaesthesia unit are very reduced and allow its installation also in
small rooms or small working areas or combined with pendant lifting systems. Two handles on the
lateral perimeter allow to easily grasp and move the anaesthesia unit.
On the rear side, a full-height steel rod is available to fix the patient monitoring.

1.3.1 Valves group - Breathing System

Completely renewed and performing, it presents the following definite advantages.
Extraction from above and perfect integration with the work shelf.
Automatic connections with double seal gaskets against accidental leaks.
Mono-bloc, fully stérilisable in autoclave.
Calibration of flow and O
access to flow sensor and O
The CO
absorber canister is located in the upper side and it easy to disconnect by apposite
2
unlock lever. With canister inserted the system makes automatic configuration in rebreathing
modality; with taken off canister, the system makes automatic configuration in non rebreathing
modality (real open circuit). It is possible to put and take off the canister during intervention.
The canister is available in version for reuse and disposable version including soda lime
granules.
Low periodic and extraordinary maintenance operations; easy training of technical personnel
thanks to the extreme rationality of the system and to the drastic reduction of accidental leaks.
Morpheus_ND
sensors can be performed in automatic mode by the User; The
2
sensor is simple and immediate.
2