Cleaning, disinfection and sterilisation table

Component
Procedure
Outer casing
Use a moistened disposable cloth with neutral
detergent or a chemical substance or the like.
Use water to remove any remaining traces of
chemical.
The User may use disinfectants (e.g. Buraton
10 F, diluted according to the manufacturer's
instructions or VPRO 60C°) to clean the
components. Disinfectants based on the
following substances can cause damage:
Remove any dust from the surfaces or in
openings using a vacuum cleaner or a soft
cloth.
Screen
See above
Dismantle and clean: sterilize in an autoclave,
disinfect with steam or chemically.
Patient circuit
(silicone tubes)
Check that there are no splits in the tubes and
replace them if they are damaged.
Morpheus_ND
halogen-releasing compounds;
strong organic acids;
oxygen-releasing compounds.
To avoid damaging the labels and outer surfaces of the anaesthesia
unit, use only the chemical substances listed.
The patient circuit can be sterilized by means of steam but this can
lead to early wear of the tubes.
Yellowing and reduced flexibility are side effects caused by
sterilization using steam.
Notes
Make sure that no sprays or liquids
penetrate inside the equipment and
the connectors.
Do not use cloths or sponges that
could scratch the surface.
121°C rubber cycle.
Before using again, eliminate any
humidity inside the tubes by means
of compressed air.