TRACe Subsystem

Key access:
Key access:
The TRACe subsystem controls access to the internal trace
memory of the analyzer.
Transfer Trace Data
:TRACe[:DATA]? <trace_name>
This command transfers trace data from the instrument to the
controller. <trace_name> is TRACE1|2|3|4.
Example:
:TRAC:DATA? TRACE1
Remarks:
Commands :MMEM:STOR:TRAC and :MMEM:LOAD:TRAC are used
to transfer trace data to, or from, the internal drive.
Select Trace Display Mode
:TRACe1|2|3|4:MODE WRITe|MAXHold|MIN-
Hold|VIEW|BLANk
:TRACe1|2|3|4:MODE?
Selects the display mode for the selected trace.
Write puts the trace in the normal mode, updating the data.
Maximum hold displays the highest measured trace value for all
the data that has been measured since the function was turned on.
Minimum hold displays the lowest measured trace value for all the
data that has been measured since the function was turned on.
View turns on the trace data so that it can be viewed on the
display.
Blank turns off the trace data so that it is not viewed on the display.
> Clear Write|Max Hold|Min Hold|View|Blank
View/
Trace
Trace Math Off
:TRACe:MATH:OFF
Turn off all the trace match function.
> More > Trace math > A-B->C Off and A+B->C off
View/
Trace
Command Reference
TRACe Subsystem
163