Ieee Common Commands - Keysight N9320B Programmer's Manual

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Command Reference

IEEE Common Commands

IEEE Common Commands
60
The first few pages of this chapter contain common commands
specified in IEEE Standard 488.2-1992, IEEE Standard Codes,
Formats, Protocols and Common Commands for Use with
ANSI/IEEE Std 488.1-1987. New York, NY, 1992.
Following these commands, the Keysight N9320B spectrum
analyzers SCPI commands are listed.
Clear Status
*CLS
Clears the status byte register. It does this by emptying the error
queue and clearing all bits in all of the event registers. The status
byte register summarizes the states of the other registers. It is also
responsible for generating service requests.
Remark:
See *STB?
Standard Event Status Enable
*ESE <number>
*ESE?
Sets the bits in the standard event status enable register. This
register monitors I/O errors and synchronization conditions such as
operation complete, request control, query error, device dependent
error, execution error, command error and power on. A summary
bit is generated on execution of the command.
The query returns the state of the standard event status enable
register.
Range:
Integer, 0 to 255
Example:
*ESE 36 Enables the Standard Event Status Register to monitor
query and command errors (bits 2 and 5).
*ESE? Returns a 36 indicating that the query and command status
bits are enabled.

Hide quick links: