Command Reference

READ Subsystem

READ Subsystem
118
Does not preset the measurement to the factory default settings.
For example, if you have previously initiated the ACP measurement
and you send READ:ACP? it will initiate a new measurement using
the same instrument settings.
• Initiates the measurement and puts valid data into the output
buffer. If a measurement other than the current one is specified,
the instrument will switch to that measurement before it initiates
the measurement and returns results.
For example, suppose you have previously initiated the ACP
measurement, but now you are running the channel power
measurement. Then you send READ:ACP? It will change from
channel power back to ACP and, using the previous ACP settings,
will initiate the measurement and return results.
• Blocks other SCPI communication, waiting until the measurement
is complete before returning the results
ACP measurement
:READ:ACPower:LOWer?
Return the value of lower-adjacent channel power (dBc).
:READ:ACPower:UPPer?
Return the value of upper-adjacent channel power (dBc).
:READ:ACPower:MAIN?
Return the value of main channel power.
Remarks
The main channel power is returned in the current amplitude units,
and the lower and upper channel results are always returned in dB.
> ACP
Key access
Meas