Source Subsystem - Keysight N9320B Programmer's Manual

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

SOURce Subsystem

Default Unit:
Key access:
Default Unit:
Key access:
The SOURce subsystem controls the signal characteristics of the
tracking generator. Refer also to the "OUTPut Subsystem" on
page 117 which contains a command that controls the tracking
generator output.
Sets the Output Power Offset Correction
:SOURce:CORRection:OFFSet <rel_ampl>
:SOURce:CORRection:OFFSet?
Specifies an offset for the displayed output power level. An offset
power level can be added to the displayed level to compensate for
system losses (for example, cable loss) or gains (for example,
preamplifier gain.) This offset does not change the power out of the
source, it only changes the display so that it reads out the actual
power delivered to the device under test.
*RST:
0 dB
Range:
–327.6 dB to 327.6 dB
Currently selected source power units
> Tracking Generator > Amptd Offset
MODE
Source Attenuation
:SOURce:POWer:ATTenuation <ampl>
:SOURce:POWer:ATTenuation?
Attenuates the source output level. Specifically setting
:SOURce:POWer:ATTenuation <ampl>
(
:SOURce:POWer:ATTenuation:AUTO OFF
*RST:
18 dB
Range:
0 dB to 60 dB in 10 dB steps
dB
> Tracking Generator > Attenuation Auto Man
MODE
Command Reference
SOURce Subsystem
sets the mode to manual
).
153

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents