Keysight N9320B Programmer's Manual page 147

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Command Reference
SENSe Subsystem
Set the Abs Stop
[:SENSe]:SEMask:OFFSet:LIST:STARt:ABSolute
<ampl>,...[<ampl>] (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:STARt:ABSolute?
[:SENSe]:SEMask:OFFSet:LIST:STOP:ABSolute:COUPle
OFF|ON|0|1,...OFF|ON|0|1 (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:STOP:ABSolute:COUPle?
Enables you to enter an absolute level limit at Stop Freq for the
currently offset ranging from –200.00 to +50.00 dBm with 0.01 dB
resolution, and to toggle this function between Couple and Man. If
set to Couple, this is coupled to Abs Start to make a flat limit line. If
set to Man, Abs Start and Abs Stop you can enter different values
to make a sloped limit line.
*RST:
Dependent upon Offset selected.
Range:
–200 dBm to 50 dBm
Example:
SEM:OFFS:LIST:STOP:ABS –20 dBm
SEM:OFFS:LIST:STOP:ABS:COUP 0
> More > Spectrum Emission Mask > Offset/Limits > More >
Key access:
Meas
Abs Stop
Set the Rel Stop
[:SENSe]:SEMask:OFFSet:LIST:STARt:RCARrier
<ampl>,...[<ampl>] (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:STARt:RCARrier?
[:SENSe]:SEMask:OFFSet:LIST:STOP:RCARrier:COUPle
OFF|ON|0|1,...OFF|ON|0|1 (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:STOP:RCARrier:COUPle?
Enables you to enter an absolute level limit at Stop Freq for the
currently offset ranging from –150.00 to +50.00 dB with 0.01 dB
resolution, and to toggle this function between Couple and Man. If
set to Couple, this is coupled to Rel Start to make a flat limit line. If
set to Man, Abs Start and Rel Stop you can enter different values to
make a sloped limit line.