Keysight N9320B Programmer's Manual page 99

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Carrier Frequency Offset Upper Limit
DANalyse:AM:LIMit:FOFFset:UPPer <freq>
DANalyse:AM:LIMit:FOFFset:UPPer?
Specifies the carrier frequency offset upper limit for the AM signal.
*RST:
10 kHz
> AM Modulation Analysis > More > Limit > CarrFreOff Up
Key access:
MODE
X Scale/Division
DANalyse:AM|FM:VIEW:X:PDIVision <time>
DANalyse:AM|FM:VIEW:X:PDIVision?
Specifies the time for every division of X scale.
> AM/FM Modulation Analysis > X Scale > Scale/DIV
Key access:
MODE
X Scale Reference Value
DANalyse:AM|FM:VIEW:X:RVALue <time>
DANalyse:AM|FM:VIEW:X:RVALue?
Specifies the reference value of the X scale.
> AM/FM Modulation Analysis > X Scale > Ref Value
Key access:
MODE
X Scale Reference Position
DANalyse:AM|FM:VIEW:X:RPOSition LEFT|CENTer|RIGHt
DANalyse:AM|FM:VIEW:X:RPOSition?
Sets the reference position of the X scale.
> AM/FM Modulation Analysis > X Scale > Ref Position
Key access:
MODE
X Auto Scale
DANalyse:AM|FM:VIEW:X:AUTO OFF|ON|1|0
DANalyse:AM|FM:VIEW:X:AUTO?
Sets the X scale according to the test result dynamically.
> AM/FM Modulation Analysis > X Scale > Auto Scale
Key access:
MODE
Command Reference
DANalyse Subsystem
93

Hide quick links:

Advertisement

Table of Contents
loading