Command Reference
SENSe Subsystem
Key access:

[:SENSe]:FREQuency Subsection

Key access:
148
The method is controlled by
• Negative peak detection displays the lowest sample taken
during the interval being displayed.
• Positive peak detection displays the highest sample taken
during the interval being displayed.
• Sample detection displays the sample taken during the interval
being displayed, and is used primarily to display noise or
noise-like signals. In sample mode, the instantaneous signal
value at the present display point is placed into memory. This
detection should not be used to make the most accurate
amplitude measurement of non noise-like signals.
• Average detection is used when measuring the average value of
the amplitude across each trace interval (bucket). The averaging
method used by the average detector is set to either video or
power as appropriate when the average type is auto coupled.
• Normal detection selects the maximum and minimum video
signal values alternately. When selecting Normal detection,
"Norm" appears in the upper-left corner.
*PST:
Positive
> Detector
Det/
Display
Center Frequency
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer UP|DOWN
[:SENSe]:FREQuency:CENTer?
Set the center frequency.
*RST:
1.500 GHz
Range:
–80 MHz through 3.08 GHz
> Center Freq