Standard Event Status Register - Keysight N9320B Programmer's Manual

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Status Registers
Status Register System

Standard Event Status Register

32
The standard event status register is used to determine the specific
event that sets bit 5 in the status byte register. The standard event
status register does not have negative and positive transition
registers, nor a condition register. Use the IEEE common
commands to access the register.
To query the standard event status register, send the *ESR
command. The response will be the decimal sum of the bits which
are set to 1. For example, if bit number 7 and bit number 3 are set
to 1, the decimal sum of the 2 bits is 128 plus 8. So the decimal
value 136 is returned.
Figure 2
Standard Event Status Register Diagram
0 1 2
3 4 5 6 7
&
&
&
&
&
&
0 1 2 3 4 5 6 7
Operation Complete
Request Bus Control
Query Error
Device Dependent Error
Execution Error
Command Error
User Request
Power On
Standard Event
Status Register
&
&
Event Enable Register
+
To Status Byte Register bit #5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents