Calculate:marker Subsection - Keysight N9320B Programmer's Manual

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Command Reference
CALCulate Subsystem

CALCulate:MARKer Subsection

Key access:
N O T E
Key access:
Default Unit:
Key access:
70
Markers All Off on All Traces
:CALCulate:MARKer:AOFF
Turns off all markers on all the traces.
> All Off
Marker
Continuous Peaking Marker Function
:CALCulate:MARKer[n]:CPEak[:STATe] OFF|ON|0|1
:CALCulate:MARKer[n]:CPEak[:STATe]?
Turns on or off continuous peaking. It continuously puts the
selected marker on the highest displayed signal peak.
This function is intended to maintain the marker on signals with a
frequency that is changing, and an amplitude that is not changing.
*RST:
Off
Remarks:
This command may not be used to activate a given marker.
> More > Continuous Pk On Off
Peak
Search
Frequency Counter Marker Resolution
:CALCulate:MARKer:FCOunt:RESolution <real>
:CALCulate:MARKer:FCOunt:RESolution?
Sets the resolution of the marker frequency counter. Setting the
resolution to AUTO will couple the marker counter resolution to the
frequency span.
*RST:
1 kHz
Range:
0.1 Hz to 1 kHz
Hz
> Function > Freq Counter > Resolution Auto Man
Marker