Keysight N9320B Programmer's Manual page 85

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Marker to Trace Auto
:CALCulate:MARKer[n]:TRACe:AUTO OFF|ON|0|1
:CALCulate:MARKer[n]:TRACe:AUTO?
Turns on or off the automatic marker to trace function.
*RST:
AUTO
> More > Marker Trace Auto 1 2 3 4
Key access:
Marker
Continuous Signal Tracking Function
:CALCulate:MARKer[n]:TRCKing[:STATe] OFF|ON|0|1
:CALCulate:MARKer[n]:TRCKing[:STATe]?
Turns on or off marker signal tracking. It continuously puts the
selected marker on the highest displayed signal peak and moves it
to the center frequency. This allows you to keep a signal that is
drifting in frequency, on the display.
*RST:
Off
Remarks:
When a measurement under the front panel MEAS key is started,
this command is turned off. If this command is turned on when any
of the MEAS key measurements are in progress, that measurement
will be stopped.
Key access:
Frequenc
Marker X Value
:CALCulate:MARKer[n]:X <param>
:CALCulate:MARKer[n]:X?
Position the designated marker on its assigned trace at the
specified trace X value. The value is in the X-axis units (which is
often frequency or time).
The query returns the X value of the designated marker.
Key access:
Marker
> Signal Track On Off
Command Reference
CALCulate Subsystem
79

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents