Keysight N9320B Programmer's Manual page 139

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Set the Occ BW% Pwr
[:SENSe]:OBWidth:PERCent <percent>
[:SENSe]:OBWidth:PERCent?
Enables you to change the percentage of signal power used when
determining the occupied bandwidth.
*RST:
99.0%
Range:
10% - 99.99%
Example:
OBW:PERC 50
Key access:
BW/
Avg
Set the X dB value
[:SENSe]:OBWidth:XDB <dB value>
[:SENSe]:OBWidth:XDB?
Enables you to specify the power level used to determine the
emission bandwidth as the number of dB down from the highest
signal point (P1), within the occupied bandwidth span. This
function is an independent calculation from the OBW calculation.
The x dB Bandwidth result is also called the emissions bandwidth,
or EBW.
*RST:
–26 dB
Range:
–100.0 dB through –0.1 dB
Example:
OBW:XDB –50 dB
Key access:
BW/
Avg
> Occupied BW > Occ BW% Pwr
> Occupied BW > Occ BW% Pwr
Command Reference
SENSe Subsystem
133

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents