Calculate:ntdata Subsection - Keysight N9320B Programmer's Manual

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Span Markers Span X Value
:CALCulate:MARKer[n]:X:SPAN <param>
:CALCulate:MARKer[n]:X:SPAN?
Change the frequency span of the designated span-type marker
pair to position the markers at the desired trace X values. The value
is in the X-axis units (which is usually frequency or time).
The query returns the current X value frequency span of the
designated markers. If span markers are not selected, the query
returns the latest marker reading as a span (always positive).
Key access:
Marker
Marker Read Y Value
:CALCulate:MARKer[n]:Y?
Reads the current Y value for the designated marker or delta on its
assigned trace. The value is in the Y-axis units for the current trace
(which is often dBm).
Remarks:
This command can be used to read the results of marker functions
such as and noise that are displayed in the marker value field on
the analyzer.

CALCulate:NTData Subsection

Normalize the Trace Data
:CALCulate:NTData[:STATe] OFF|ON|0|1
:CALCulate:NTData[:STATe]?
One sweep of trace data is copied to trace 3, which is used as the
reference trace. Then for all subsequent trace sweeps, display
trace 1 = data collected into trace 1 – data in trace 3.
Key access:
MODE
> Delta > Span Pair
> Tracking Generator > More > Normalize > Normalize On Off
Command Reference
CALCulate Subsystem
81

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents