Keysight N9320B Programmer's Manual page 167

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Power on Type
:SYSTem:PON:TYPE PRESet|LAST
:SYSTem:PON:TYPE?
Returns the instrument to a set of defined conditions. The
particular set is selected by :SYSTem:PRESet:TYPE. This command
does not change any persistent parameters. The term persistent
means that the command retains the setting previously selected,
even through a power cycle.
Key access:
Preset/
System
Preset
:SYSTem:PRESet
Returns the instrument to a set of defined conditions. The
particular set is selected by :SYSTem:PRESet:TYPE. This command
does not change any persistent parameters. The term persistent
means that the command retains the setting previously selected,
even through a power cycle.
Key access:
Preset/
System
Persistent State Reset
:SYSTem:PRESet:PERSistent
Sets the persistent state values to their factory defaults. The term
persistent means that the command retains the setting previously
selected, even through a power cycle. Examples of persistent
functions are: power-on type, and preset type.
Save User Preset
:SYSTem:PRESet[:USER]:SAVE
Saves the current instrument conditions as the user preset
condition.
Key access:
Preset/
System
> Pwr On/Preset > Power On Last Preset
> Preset
> Power On/Preset > Save Type Preset
Command Reference
SYSTem Subsystem
161

Hide quick links:

Advertisement

Table of Contents
loading