[:Sense]:Chpower Subsection - Keysight N9320B Programmer's Manual

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Set the Measurement Type
[:SENSe]:ACPower:TYPE TPRef|PSDRef
[:SENSe]:ACPower:TYPE?
Press Meas Type to specify the reference for the measurement,
either Total Pwr Ref or PSD Ref. Relative values can be displayed
referenced to either the total power (Total Pwr Ref) or the power
spectral density (PSD Ref) measured in the main channel.
*RST:
Total Pwr Ref
Example:
ACP:TYPE PSDR
Key access:
Meas

[:SENSe]:CHPower Subsection

Set the Average
[:SENSe]:CHPower:AVERage:COUNt <integer>
[:SENSe]:CHPower:AVERage:COUNt?
[:SENSe]:CHPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:CHPower:AVERage[:STATe]?
To specify the number of measurement averages used when
calculating
the measurement result set Avg Number to ON. The average will
be displayed at the end of each sweep. Setting Avg Number to OFF
disables measurement averaging.
*RST:
10/Off
Range:
1 to 1000
Example:
CHP:AVER:COUN 20
CHP:AVER OFF
Key access:
Meas
> ACP > Meas Type
> Channel Power > Avg Number On Off
Command Reference
SENSe Subsystem
127