Keysight N9320B Programmer's Manual page 75

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Display the Limit Margin
:CALCulate:LLINe[1]|2:MARGin:STATe OFF|ON|0|1
:CALCulate:LLINe[1]|2:MARGin:STATe?
Displays a measurement margin that is added to the designated
limit line to do secondary testing of the data.
*RST:
Off
Key access:
>
Det/
Display
Control Limit Line Testing
:CALCulate:LLINe[1]|2:STATe OFF|ON|0|1
:CALCulate:LLINe[1]|2:STATe?
Turns limit line testing on/off. The limit and margin will only be
tested if they are displayed. Use
on the display of limit lines, and
to turn on the display of margins. If margin and limit display are
both turned off, limit test is automatically turned off. Use
:CALCulate:LLINe[1]|2:FAIL?
limit line state has been turned on.
*RST:
Off
Key access:
>
Det/
Display
Select the Type of Limit Line
:CALCulate:LLINe[1]|2:TYPE UPPer|LOWer
:CALCulate:LLINe[1]|2:TYPE?
Sets a limit line to be either an upper or lower type limit line. An
upper line will be used as the maximum allowable value when
comparing with the data.
*RST:
Lower
Remarks:
If a margin has already been set for this limit line, and this
command is used to change the limit type, then the margin value is
reset to 0 dB.
>
Key access:
Det/
Display
Limits > Limit 1|2 > Margin On Off
to return the state of pass or fail after
Limits > Limit 1|2 > Limit On Off
Limits > Limit 1|2 > Type Upper Lower
Command Reference
CALCulate Subsystem
:CALCulate:LLINe[1]|2:DISPlay
:CALCulate:LLINe[1]|2:MARGin:STATe
to turn