individual limits when initially set.
For example, if the reference level is (first) set to –20 dBm, then the
reference level offset can be set to values of –307.6 dB to 327.6 dB.
In the case of a 327.6 dB reference level offset, the resultant
reference level value changes to 307.6 dBm. The reference level
value range can be initially set to values from –100 to 50 dBm.
Vertical Axis Scaling
:DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing LINear|LOGa-
rithmic
:DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing?
Specifies the vertical graticule divisions as log or linear units.
*RST:
Logarithmic
Key access:
Amplitud
Active function Position
:DISPlay:AFUnction:POSition CENTter|TOP|BOTTom
:DISPlay:AFUnction:POSition?
Selects the screen position for the Active Function Display.
Depending on the type of trace data being viewed, you can move
the active function display position to minimize its visual
interference with your screen data.
*RST:
Center
Key access:
Det/
Display
> Scale Type
> Active Fctn Position
Command Reference
DISPlay Subsystem