Keysight N9320B Programmer's Manual page 157

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Input Port Attenuator Auto
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Select the input port attenuator range to be set either
automatically (On) by the Reference Level Setting or manually
(Off).
*RST:
On
Key access:
Amplitud
Attenuation Limit
[:SENSe]:POWer[:RF]:ATTLimit:STATe OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTLimit:STATe?
Toggles the attenuation limit function between On and Off.
*RST:
On
Key access:
Amplitud
Input Port Power Gain
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Turns the internal preamp on or off.
*RST:
Off
Remarks:
This command is only available with Option PA3 installed.
Key access:
Amplitud
> Attenuation
> More > Att Limit
> Int Preamp On Off
Command Reference
SENSe Subsystem
151

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents