offset value results in the display of trace data prior to the trigger
event. A positive trigger offset value results in an effective delay in
the display of trace data after the trigger event.
The trigger offset value used when the feature is enable will
depend on the following parameters:
• Normal trigger offset value originally entered
• Specific instrument hardware in use
• Sweep time
Key access:
Sweep/
Trig
Turn Trigger Offset On/Off
:TRIGger[:SEQuence]:OFFSet:STATe ON|OFF
:TRIGger[:SEQuence]:OFFSet:STATe?
This command turns the trigger offset on/off.
*RST:
Off
Key access:
Sweep/
Trig
Trigger Source
:TRIGger[:SEQuence]:SOURce IMMediate|VIDeo|EXTernal
:TRIGger[:SEQuence]:SOURce?
Specifies the source (or type) of triggering used to start a
measurement.
Immediate is free-run triggering
Video triggers on the video signal level
External allows you to connect an external trigger source
*RST:
Immediate (free-run triggering)
Remarks:
Free-run activates the trigger condition that allows the next sweep
to start as soon as possible after the last sweep.
Key access:
Sweep/
Trig
> More > More > Trig Offset On Off
> More > More > Trig Offset On Off
> More > Free Run|Video|Line|External Pos Neg
Command Reference
TRIGger Subsystem
167