Keysight N9320B Programmer's Manual page 148

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Command Reference
SENSe Subsystem
Key access:
Key access:
Key access:
142
*RST:
Dependent upon Offset selected, please refer to the User's Guide
for more details.
Range:
–150 dB to 50 dB
Example:
SEM:OFFS:LIST:STOP:RCAR
SEM:OFFS:LIST:STOP:RCAR:COUP 0
> More > Spectrum Emission Mask > Offset/Limits > More >
Meas
Rel Stop
Set the Fail Mask
[:SENSe]:SEMask:OFFSet:LIST:TEST ABSolute|AND|OR|REL-
ative,ABSolute|AND|OR|RELative,ABSolute|AND|OR|RELa-
tive,ABSolute|AND|OR|RELative
[:SENSe]:SEMask:OFFSet[n]:LIST:TEST?
Displays the menu to select one of the following logic keys for fail
conditions between the measurement results and the test limits:
Absolute, Relative, Abs AND Rel, and Abs OR Rel.
*RST:
Dependent upon Offset selected, please refer to the User's Guide.
Example:
SEM:OFFS:LIST:TEST ABS,REL,ABS AND REL,ABS OR REL
>More >Spectrum Emission Mask >Offset/Limits >More
Meas
>More >Fail Mask
Set the Sweep Time
[:SENSe]:SEMask:SWEeptime <number>
[:SENSe]:SEMask:SWEeptime?
[:SENSe]:SEMask:SWEeptime:AUTO OFF|ON|0|1
[:SENSe]:SEMask:SWEeptime:AUTO?
Specifies the sweep time used in measuring the power in the main
channel.
*RST:
Auto
Range:
8.37 ms through 4 ks
Example:
SEM:SWE 4 s
> More> Spectrum Emission Mask > Ref Channel > Sweep Time
Meas
20 dB

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents