Keysight N9320B Programmer's Manual page 79

Spectrum analyzer
Hide thumbs Also See for N9320B:
Table of Contents

Advertisement

Marker Peak (Minimum) Search
:CALCulate:MARKer[n]:MINimum
Places the selected marker on the lowest point on the trace that is
assigned to that particular marker number.
Key access:
Peak
Search
Marker Mode
:CALCulate:MARKer[n]:MODE POSition|DELTa
:CALCulate:MARKer[n]:MODE?
Selects the type of markers that you want to activate.
Position selects a normal marker that can be positioned on a trace
and from
Delta activates a pair of markers, one of which is fixed at the
current marker location. The other marker can then be moved
around on the trace. The marker readout shows the difference
between the two markers.
Remarks:
If a marker is not active when the mode is queried, "Off" will be
returned.
> Normal
Key access:
Marker
> Delta > Delta
Marker
Define Peak Excursion
:CALCulate:MARKer:PEAK:EXCursion <rel_ampl>
:CALCulate:MARKer:PEAK:EXCursion?
Specifies the minimum signal excursion above the threshold for the
internal peak identification routine to recognize a signal as a peak.
This applies to all traces. The excursion is the delta power from the
noise level to the signal peak.
:CALCulate:MARKer:PEAK:SEARch:MODE.
*RST:
6 dB
Range:
0 to 100 dB
Key access:
Peak
Search
> Min Search
which trace information will be generated.
> More > Search Criteria > Peak Excursion
Command Reference
CALCulate Subsystem
See
73

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents