SENSe Subsystem

[:SENSe]:SEMask Subsection

Key access:
Key access:
136
Set the Average
[:SENSe]:SEMask:AVERage:COUNt <integer>
[:SENSe]:SEMask:AVERage:COUNt?
[:SENSe]:SEMask:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:SEMask:AVERage[:STATe]?
To specify the number of measurement averages used when
calculating the measurement result set Avg Number to ON. The
average will be displayed at the end of each sweep.
*RST:
10/Off
Range:
1 to 1000
Example:
OBW:SEM COUN 20
> More > Spectrum Emission Mask > Avg Number On Off
Meas
Set the Chan Integ BW
[:SENSe]:SEMask:BANDwidth|BWIDth:INTegration <freq>
[:SENSe]:SEMask:BANDwidth|BWIDth:INTegration?
Specifies the integration bandwidth used in calculating the power
in the main channel.
*RST:
3.84 MHz
Range:
10% to 100% of the setting of Chan Span
Example:
SEM:BAND:INT 4 MHz
>More>Spectrum Emission Mask>Ref Channel>Chan Integ BW
Meas

Hide quick links: