Basic Operation
Chapter 3

3.1 Preparing the Flat Panel Sensor

This section describes how to prepare the fl at panel sensor.

3.1.1 Type of Flat Panel Sensor

DR-ID1271SE, DR-ID1272SE, DR-ID1273SE, DR-ID1274SE, DR-ID1275SE
The battery pack (optional) is required when the fl at panel sensor is used in wireless communication
mode.

3.1.2 Number of the Connectable Flat Panel Sensors

To enable the fl at panel sensor, its ID needs to be registered in advance by our offi cial dealer or
FUJIFILM Representative.
Up to a hundred fl at panel sensors can be registered.
Up to fi ve fl at panel sensors can be connected. Up to two fl at panel sensors can be connected to
one power supply unit in wired communication mode. Up to one power supply units can be used.
3.1.3 Connecting/Disconnecting the Flat Panel Sensor
Connector
1
Disconnect the connector.
Press the latches on both sides of the connector.
Do not place the connecter on the fl oor.
When multiple fl at panel sensors are used, make sure that the READY lamp among the status lamps of the fl at
panel sensor to be used for an exposure is lit.
Connect/Disconnect the connector straight to the fl at panel sensor. If connected/disconnected at an angle, the
connector may be damaged.
2
Connect the connector.
Press the connector into the insertion section.
Make sure that the latches on both sides are
properly engaged when connecting the connector.
If the connector is inserted incompletely, the power
may turn off.
DR-ID1270 Operation Manual
3
3-1