Table B-3. Alarm Trigger Values - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

C-Link Protocol Commands
Diagnostics
Diagnostics
B-18
Model 42i Instruction Manual
Send:
set alarm trig conc no 1
Receive:
set alarm trig conc no 1 ok
Table B–3. Alarm Trigger Values
Alarm Trigger
Value
00
Floor
01
Ceiling
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt mb
Receive:
diag volt mb 24.1 14.9 4.9 3.2 -3.2
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, positive 3.3, and positive 15. Each
voltage value is separated by a space.
Send:
diag volt mib
Receive:
diag volt mib 24.1 14.9 -14.9 4.9 3.2 14.9
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt iob
Receive:
diag volt iob 24.1 4.9 3.2 -3.2
version
This command reports the version of all the firmware components. The
following example shows a list of the firmware components that were
displayed by issuing the version command.
Send:
version
Receive:
version
Program = 01.05.79.225
Library = 01.01.60.167
Kernal = 2.4.24-uc0-003-Thermo
Board = 81, File = /usr/application.hex
Board App = 11.3.100 BI 4.0.97
File App = 11.3.100 BI 4.0.97