Timed Interval; Calibration Due Alarm - Thermo Scientific Eutech EC 1720 User Manual

Conductivity meter
Table of Contents

Advertisement

3.2.11 Timed Interval

When Timed is set as the Read Type, set the time interval
from 5 seconds to 60 minutes. This time interval is used to
automatically save and export measurements.
Press the  or  key to scroll through the Setup
Menu list until the Timed Interval item is shown.
Press the  or  key to adjust the time value.
a. Press the  key to increase the time value by
one second. Press and hold the  key to
quickly increase.
b. Press the  key to decrease the time value by
one second. Press and hold the  key to
quickly decrease.
Press the "Enter" key to save the change.
20
Chapter 3 | Setup Menu

3.2.12 Calibration Due Alarm

Set the calibration due alarm interval from 0 hours (off) to
168 hours. An alarm is triggered if a calibration is not
performed within the specified time interval.
Press the  or  key to scroll through the Setup
Menu list until the Calibration Due Alarm item is
shown.
Press the  or  key to adjust the alarm value.
a. Press the  key to increase the alarm value by
one hour. Press and hold the  key to quickly
increase.
b. Press the  key to decrease the alarm value
by one hour. Press and hold the  key to
quickly decrease.
Press the "Enter" key to save the change.