Table B-3. Alarm Trigger Values; Diagnostics - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Diagnostics

Datalogging
Thermo Fisher Scientific
These commands set the NO, NO
alarm minimum value, where value is set to either floor or ceiling,
according to
Table
B–3. The example below sets the NO concentration
minimum alarm trigger to ceiling.
Send:
set alarm trig conc no 1
Receive:
set alarm trig conc no 1 ok
Table B–3. Alarm Trigger Values
Value
Alarm Trigger
00
Floor
01
Ceiling
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt mb
Receive:
diag volt mb 24.1 14.9 4.9 3.2 -3.2
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, positive 3.3, positive 18 IR, positive 18
MOT, and VBIAS. Each voltage value is separated by a space.
Send:
diag volt mib
Receive:
diag volt mib 24.1 14.9 -14.9 4.9 3.2 17.9 17.9
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt iob
Receive:
diag volt iob 24.1 4.9 3.2 -3.2
clr records
This command will clear all long and short records that have been saved.
Send:
clear records
Receive:
clear records ok
C-Link Protocol Commands
, NO
and prereactor concentration
2
x,
Model 42i Trace Level Instruction Manual
Diagnostics
B-17