8. Quality Control
8.3.4
Fluorescence Polarization (infinite
Precision
Measurement Parameters:
138
®
For the infinite
F200 with the option 'Fluorescence Polarization' the following tests
may be performed to prove the specifications:
•
Precision/Repeatability
These test instructions are valid for the infinite
•
Standard version
•
Spectrally enhanced version
Perform the following measurement to determine the precision:
Parameter
Reading Mode
Exc Wavelength
Em Wavelength
Number of flashes
Integration Time
Lag Time
Time between Move and Read
Gain
Plate Type
Reference from/to
Reference blank from/to
Reference value
Measurement blank from/to
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
®
F200 only)
®
F200:
Setting
Fluorescence Polarization
485 (20 nm)
535 (25 nm)
25
40
0
0
Optimal
GRE96fb
A1 – D1
A2 – D2
20 mP
same as reference blank
2008-07
Need help?
Do you have a question about the infinite 200 and is the answer not in the manual?