2. General Description
Injection Mode
40
The injection settings can be adjusted via the software:
Injection
Select Injector:
Injector A or Injector B can be selected. It is not possible to select
both injectors on one strip. If a measurement with two injectors is to
be performed, two injector strips are necessary.
Speed: The injection speed is selectable from 100 – 300 µl/sec for
each injector.
Select a 'Refill speed' from 100 – 300 µl/sec. for each injector or
check the 'Refill Speed equal to Injection Speed' box.
Select refill mode 'Standard' if refill should be performed when
syringe is empty (multiple injection steps are performed before
refilling, refill occurs after dispensing approx. 800 µl). Select 'Refill
for every injection' if refill should be performed for every injection
step.
Injection
The injection volume depends on the microplate type. The plate
volume
definition files include a so-called working volume. This working
volume defines the maximum volume to be injected into the
selected microplate. Therefore, always make sure that the selected
plate definition file contains a correct setting for the working volume.
The maximum injection volume is 800 µl/injection strip. If volumes
greater than 800 µl are to be injected (e.g. into 6-well plates), more
than one injection strip has to be used.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4