Absorbance Baseline Flatness - tecan infinite 200 Instructions For Use Manual

Table of Contents

Advertisement

8.3.8
Absorbance Baseline Flatness (infinite
Measurement Parameters:
Plate Layout:
Material/Reagents:
Calculation of Baseline Flatness:
2008-07
Perform the following measurement to determine the baseline-flatness:
Parameter
Reading Mode
Measurement Wavelength from/to
Number of flashes
Settle Time
Plate Type
Part of the Plate
No plate is necessary for measurement – the plate carrier has to be empty for this
measurement.
No material or reagents necessary for this test.
Calculate the standard deviation.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
8. Quality Control
®
M200)
Setting
Absorbance
300 - 700 nm
25
0
GRE96ft
A1
143

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the infinite 200 and is the answer not in the manual?

Table of Contents

Save PDF