4. Defining Filter Slides (infinite® F200 only)
STOP
62
Fluorescence polarization mode on Position 1 requires the same filter
settings on Position 2 and vice versa. Fluorescence polarization mode
on Position 3 requires the same filter settings on Position 4 and vice
versa. This is performed automatically.
Make sure that the filter slide contains polarizers together with the filters
defined for fluorescence polarization.
Wavelength: Enter the filter wavelength within the following range:
(1) Fluorescence intensity mode: 230 to 850 nm (Excitation) and 280 to 850 nm
(Emission)
(2) Fluorescence polarization: 300 to 850 nm (Excitation) and 330 to 850 nm
(Emission)
(3) Absorbance mode: 230 to 1000 nm
Bandwidth: Enter the bandwidth (nm) of the filter
(4) Accept the new filter values by clicking Save. By closing the Filter Definition
dialog the system is ready to collect data with the new filters.
Description: This field can be used for individual user's remarks about the filter,
e.g. filter name, application, etc.
No special characters (blank, ?, $, %, ., /, etc.) except '_' are allowed for
Purchase Date: This option enables the user to enter the purchase or installation
date of the filter
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Note
Caution
Note
the filter slide description.
2008-07