10. Troubleshooting

Error #
Error Text
1
Command is not valid
2
Parameter out of range
3
Wrong number of parameters
4
Invalid parameter
5
Invalid Parameter at pos
6
[prefix] is missing
7
RS485 Timeout at module
[module descr]
8
Invalid module number [Nr]
9
Binary Transfer command:
[cmd] at module [n]
10
Error at command [cmd] at
module [n],
11
LID open
12
LUMI FIBER broken
13
Z Motor out of Safety-Range
14
Filter is not defined
15
X drive init error
16
Y drive init error
2008-07
Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
Unspecific internal communication error. Please report this
error to your local Tecan customer support office.
Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
Unspecific internal communication error. Please report this
error to your local Tecan customer support office.
Unspecific internal communication error. Please report this
error to your local Tecan customer support office.
Plate transport or filter slide lid were open during a
measurement or the instrument was used in very bright
environment (<< 500 LUX). Please check if the lid closes
completely or if the environment was too bright.
Hardware Failure Luminescence Module. Please report this
error to your local Tecan customer support office.
Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
Hardware Failure Plate Transport Module. Please report this
error to your local Tecan customer support office.
Hardware Failure Plate Transport Module. Please report this
error to your local Tecan customer support office.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
10. Troubleshooting
Description
155