2. General Description
STOP
STOP
26
A prime volume below 2 ml in an empty system may result in incomplete
filling of the system, and therefore may negatively affect assay
Figure 2-5 'Service Position' of the injectors. The injectors are removed from the
carrier slot and inserted into the holder of the injector carrier system.
Do not touch the injector needles. They can become easily bent or
misaligned, which can cause injection problems or damage the
If the injector carrier is not inserted correctly in the injector port, the
injector sensor does not detect the inserted injector and therefore
washing and priming is enabled which can damage the instrument. In
addition to this, the actions 'dispense' and 'inject' will not be possible.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Caution
performance.
Caution
instrument.
Service-Position
2008-07