tecan infinite 200 Instructions For Use Manual page 44

Table of Contents

Advertisement

2. General Description
Example 2: Enliten
Detection Kit for ATP (Promega Corp.)
Plate
Part of the plate
Well
Injection
Wait (Timer)
Luminescence
Move Plate
44
®
ATP Assay System Bioluminescence
For assay details please refer to www.promega.com.
Select an appropriate plate type. For luminescence
measurements, white microplates are recommended. For
this example, a white 96 well plate was selected.
(Not shown); can be optionally selected if only part of the
plate should be processed
Mandatory for measurements with 'injection'
Injector A injects 100 µl with speed 100 µl/sec., refill mode:
standard
2 s wait time
Luminescence measurement
with 10 s integration time, attenuation 'none'
Plate is moved out after finishing all wells
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
2008-07