2. General Description
STOP
Washing
Washing (general procedure):
30
The injector carrier must be in the service position
Do not perform backflush when the injector is in the instrument!
Before the instrument is switched off, it is recommended to perform a wash
procedure to clean the injector system.
Wash procedure:
Washing can be performed by using the i-Control software or by using the
hardware buttons on the injector box.
1.
Fill the storage bottles with the appropriate wash reagents (distilled water,
70 % ethanol, ...) and insert the feeding tube(s). Make sure, that the
tube(s) reaches the bottom of the bottle.
2.
Remove the injector from the carrier slot and insert it into the service
position of the injector box.
3.
Put an empty container under the injector.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Caution
for the action 'backflush'.
2008-07