5. Optical System
86
Example 3:
Usage of 'Move Cuvette OUT' strip when measuring a combination of microplate
and cuvette:
Cuvette
Absorbance strip
(cuvette)
Move Cuvette (Out)
User Request
Move Cuvette (In)
Plate
Part of Plate
(not shown)
Absorbance strip
(microplate)
Move Plate
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Necessary for cuvette measurement
Up to 4 absorbance fixed wavelength strips are allowed.
Reference wavelength is only selectable when using one
absorbance fixed wavelength strip. 'Apply blanking' is
disabled when reference wavelength is selected.
Select the appropriate measurement parameters
(wavelength, number of flashes and settle time)
The cuvette holder is moved 'out'.
The user request interrupts the measurement and
therefore allows removing the cuvette from the cuvette
port. When confirming the request the measurement
continues.
The cuvette port is moved in.
Necessary for microplate measurements. Select an
appropriate plate type for the measurement.
Optional. Use the 'part of plate' strip if only a part of the
plate will be measured.
Up to 10 absorbance fixed wavelength strips are allowed.
Reference wavelength is only allowed on the first
absorbance strip. Reference wavelength is disabled on
absorbance strips 2 to 10.
Select the appropriate measurement parameters
(wavelength, number of flashes and settle time) for your
application.
Optional. To move the microplate automatically out of the
instrument when finishing the measurement, select the
'Move plate OUT'.