Using the Injection Strip:
Plate
Part of the plate
Well
Injection
Measurement strip
(Example
Absorbance)
2008-07
Select an appropriate plate type.
Optional;
Select the wells to be dispensed
The well strip is mandatory.
Injection is only possible with a 'well' strip. This strip ensures that
the following indented strips are performed well-wise.
Set up the injection parameters.
Only one injector can be selected per strip. If both injectors are
required or one injector will perform two injections, an additional
injection strip has to be inserted.
It is mandatory to use at least one measurement strip in
combination with the injection strip. The position of the
measurement strip(s) (before and/or after the injection strip)
depends on the application and is therefore user-selectable.
Make sure that the corresponding Working Volume value in your plate
definition file is higher than the volume used for injection.