8. Quality Control

8.2 Specifications - Passed/Failed Criteria

128
All specifications are subject to change without prior notification.
The following table gives an overview of the passed/failed criteria for the
specification test of the
Specification Microplate
Fluorescence Top Sensitivity
Fluorescence Top Uniformity
Fluorescence Top Precision
Fluorescence Bottom Sensitivity
Fluorescence Bottom Uniformity
Fluorescence Bottom Precision
Time Resolved Fluorescence
Sensitivity (F200)
Time Resolved Fluorescence Precision
(F200)
FP Precision (F200 only)
Luminescence Sensitivity Glow Type
Luminescence Sensitivity Glow Type
Luminescence Sensitivity Flash Type
Luminescence Sensitivity Flash Type
Absorbance Accuracy
Absorbance Baseline Flatness
(1 sigma)
Absorbance Wavelength Accuracy
Specification Cuvette (M200 only)
Absorbance Accuracy
Absorbance Baseline Flatness
(1 sigma)
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Note
®
.
infinite
200
Passed/Failed Criteria
< 20 pM Fluorescein
< 3 % CV
< 2 % CV
< 100 pM Fluorescein
< 3 % CV
< 2 % CV
< 3 pM
< 2 % CV
< 5 mP
< 3 fmol/well (Standard PMT)
< 20 fmol/well (Basic PMT)
< 80 amol/well
(Standard PMT)
< 500 amol/well (Basic PMT)
0 – 2 OD: ≤ ± 1 % + 10 mOD
2 – 3 OD: ≤ ± 2.5 %
< ± 10 mOD
≤ ± 1.5 nm
≤ ± 0.8 nm
Passed/Failed Criteria
0 – 2 OD: ≤ ± 1 % + 10 mOD
2 – 3 OD: ≤ ± 2.5 %
< ± 10 mOD
λ
> 315 nm;
λ
≤ 315 nm
2008-07