2. General Description

2.3 Injectors

20
®
The
can be optionally equipped with an injector module consisting of
infinite
200
one or two syringe pumps (XE-1000, Tecan Systems) located in a separate box,
which feed one or two injector needles. The injector needles are designed to
inject liquid in any SBS-conform microplate well types, in which the well-size is
equal to or larger than an SBS standard 384-well plate.
Figure 2-1: Schematic view of the injector module
There are up to two pumps available for the
Pump A feeds injector needle A
Pump B feeds injector needle B
®
The
can be equipped with one pump (pump A) or two pumps
infinite
200
(pumps A and B).
One Injector Option (one pump): An
allows injections in any SBS-conform microplate well types, in which the well-size
is equal to or larger than an SBS standard 384-well plate.
Two Injector Option (two pumps): Some applications, such as flash
luminescence reactions or dual reporter gene assays require the injection of two
independent liquids into the same well; therefore, Tecan Austria offers a two-
injector option.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
®
(see Figure 2-1 above):
infinite
200
®
equipped with one pump
infinite
200
2008-07