infinite
F200
Filter Handling
®
infinite
M200
Microplates to be
Measured
Plate Definition
Temperature Control
Plate Shaking
Parameters
Light Source
Optics
Fluorescence Detector
Luminescence Detector Low dark count photomultiplier tube
Absorbance Detector
Power Supply
Power Consumption
Physical
Outer Dimensions
Basic instrument
®
Weight infinite
F200
®
Weight infinite
M200
Outer Dimensions
Pump Module
Weight Pump Module
Environmental
Ambient Temperature
Operation
Non-operation
Relative Humidity
Operation
Over-voltage Category
Pollution Degree
Usage
Noise Level
Method of Disposal
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Characteristics
Software controlled
USB
External filter exchange
Monochromator-based wavelengths change – no filters
necessary
From 6-well to 384-well plates
(SBS standard formats)
Via scanning software
From 5 °C above ambient up to 42 °C
Linear and orbital shaking, amplitude selectable from 1
– 6 mm in 0.5 mm steps
Characteristics
High energy Xenon Flash lamp, life time: 10
Fused Silica Lenses
Standard: PMT
Spectrally enhanced: red-sensitive PMT
photon-counting electronics (Basic and Standard PMT)
Silicon photodiode
Auto-sensing: 100 – 120 V/220 – 240 V, 50-60 Hz
150VA
Width: 425 mm
Height: 253 mm
Depth: 457 mm
14 kg
15.8 kg
Width: 250 mm
Height: 155 mm
Depth: 156 mm
3.4 kg
+ 15 °C to + 30 °C
- 20 °C to + 60 °C
< 90 % non condensing
II
2
General Laboratory Instrument
< 60 dBA
Electronic waste
(infectious waste)
8
flashes
16.73 inches
9.96 inches
17.99 inches
9.84 inches
6.10 inches
6.14 inches
+ 59 °F to + 86°F
-4 °F to + 140 °F
2008-07