8.3.2
Fluorescence Bottom
Sensitivity
Measurement Parameters:
Plate Layout:
<>
1
2
A
B
C
D
E
F
G
H
2008-07
®
For the
with the option 'Fluorescence Bottom' the following tests
infinite
200
may be performed to prove the specifications:
•
Sensitivity
•
Uniformity
•
Precision/Repeatability
These test instructions are valid for the
®
•
infinite
F200
®
•
infinite
M200
•
Standard version
•
Spectrally enhanced version
Perform the following measurement to determine the detection limit for
Fluorescein:
Parameter
Reading Mode
Exc Wavelength
Em Wavelength
Number of flashes
Integration Time
Settle Time
Gain
Plate Type
3
4
5
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
®
infinite
200
Setting
Fluorescence Bottom
®
infinite
F200
®
infinite
M200
®
infinite
F200
®
infinite
M200
25
40
0
Optimal
GRE96fb
6
7
8
8. Quality Control
:
: 485 (20 nm)
: 485 nm
: 535 (25 nm)
: 535 nm
9
10
11
12
133