STOP
STOP
2008-07
Picture 2-1: Inserting the injector carrier into the injector port
The injector carrier must be in the service position
Prime and Wash must not be performed
when the injector is in the instrument!
If the injector carrier is not inserted correctly in the injector port, the
injector sensor will not detect the inserted injector and therefore
washing and priming will be enabled, which can damage the instrument.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Caution
for washing und priming.
Caution
2. General Description