8. Quality Control
8.3.7
Absorbance Wavelength Accuracy
Measurement Parameters:
Material/Reagents:
Calculation of Wavelength Accuracy:
142
The wavelength accuracy defines the deviation of the set measurement
wavelengths from the nominal wavelength. This test is only valid for the
®
.
infinite
M200
Parameter
Reading Mode
Measurement Wavelength from/to
Step Size
Number of flashes
Settle Time
Plate Type
MultiCheck plate
Please refer to the data sheet in the instructions for use for your MultiCheck plate.
Wavelength
Accuracy
Max
theoretical Maximum
t
Max
measured Maximum
m
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Setting
Absorbance Scan
300 – 850 nm
1 nm
25
0
MultiCheck plate
=
−
Max
Max
t
m
2008-07