7. Instrument Features
7.5 Absorbance
120
Parameters
Wavelength Range
Measurement Range
The following specifications are valid for the wavelength range from
300 – 700 nm for infinite
Plate type (number of wells)
Accuracy 0 – 2 OD
Accuracy 2 – 3 OD
Baseline Flatness
M200
Wavelength Accuracy
F200
Wavelength Accuracy
The specifications are valid for measurements performed with 25 flashes (reads)
per well.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Instrument Type
®
infinite
F200
®
infinite
M200
Both
®
200 series.
96
< ± (1 % + 10 mOD)
< ± 2.5 %
± 10 mOD (1 sigma)
≤ ± 1.5 nm
≤ ± 0.8 nm
Dependent on filters used
Characteristics
230 – 1000 nm
230 – 1000 nm
no filter necessary, selection in
1 nm steps possible
0 – 3 OD
λ
> 315 nm;
λ
≤ 315 nm
2008-07