Absorbance - tecan infinite 200 Instructions For Use Manual

Table of Contents

Advertisement

7. Instrument Features

7.5 Absorbance

120
Parameters
Wavelength Range
Measurement Range
The following specifications are valid for the wavelength range from
300 – 700 nm for infinite
Plate type (number of wells)
Accuracy 0 – 2 OD
Accuracy 2 – 3 OD
Baseline Flatness
M200
Wavelength Accuracy
F200
Wavelength Accuracy
The specifications are valid for measurements performed with 25 flashes (reads)
per well.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Instrument Type
®
infinite
F200
®
infinite
M200
Both
®
200 series.
96
< ± (1 % + 10 mOD)
< ± 2.5 %
± 10 mOD (1 sigma)
≤ ± 1.5 nm
≤ ± 0.8 nm
Dependent on filters used
Characteristics
230 – 1000 nm
230 – 1000 nm
no filter necessary, selection in
1 nm steps possible
0 – 3 OD
λ
> 315 nm;
λ
≤ 315 nm
2008-07