4. Defining Filter Slides (infinite® F200 only)
Flash Counter: The flash counter monitors the number of flashes through a filter.
The flash counter number provides the user only with additional information about
the filter in use. The flash counter number is saved together with other
information about the filter on the filter slide microchip.
If you replace a filter, this information will be lost unless the last filter flash number
is manually documented by the user.
For a brand new filter, set the counter to 0. For a previously used filter, enter the
last collected flash number if the number is available.
Caution
STOP
It is recommended to manually document the last flash counter number
before replacing the filter; otherwise this information will be lost.
2008-07
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4