Plate Layout:
<>
1
2
A
B
C
D
E
F
G
H
Material/Reagents:
Calculation of Precision:
2008-07
3
4
5
Filling volume: 200 µl/well
1 nm Fluorescein (in 0.01 M NaOH) (Fluorescein sodium salt, Sigma F6377)
0.01 M NaOH (=Blank) (NaOH pellets, Merck Article No. 6495 or Sigma S8045)
1 Greiner 96-well plate, black. flat bottom
200 µl Pipette + tips
The precision is calculated from the wells filled with fluorescein. The precision
corresponds to one time standard deviation of the mP values of the fluorescein
wells.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
6
7
8
8. Quality Control
9
10
11
12
139